RADIOMETER ABL800 FLEX Operator's Manual page 338

Hide thumbs Also See for ABL800 FLEX:
Table of Contents

Advertisement

11. Troubleshooting
Analyzer messages,
Messages at
manager level
(continued)
11-56
Continued
No.
Message
878
No Rinse
solution
during
calibration
879
No Cal1
solution
during
calibration
880
No Cal2
solution
during
calibration
881
LIS/HIS:
Missing
CIC Com-
munication
Option Key
882
Mixer tray
slip detected
883
Mixer error
ABL800 FLEX Operator's Manual
Interpretation
The analyzer was not
able to retrieve Rinse
solution during
calibration.
The analyzer was
unable to retrieve Cal1
solution during
calibration.
The analyzer was
unable to retrieve Cal2
solution during
calibration.
No valid Option Key
for CIC communication
is installed on the
analyzer.
A small positioning
problem of the mixer
tray has been detected
and corrected.
The sample may not
have been adequately
mixed.
Operator action
Check level in Rinse
Solution bottle. Replace
bottle as necessary.
Check level in Cal1
solution bottle. Replace
bottle as necessary.
Check level in Cal2
solution bottle. Replace
bottle as necessary.
Contact Radiometer service
representative.
- Lift the sampler tray
cover and check for
obstructions in the mixer
tray area.
- Remove any foreign
objects.
- Remount the sampler tray
cover.
Removal condition:
Alignment operation
performed without slip.
- Lift the sampler tray
cover and check for
obstructions in the mixer
tray area.
- Remove any foreign
objects.
- Remount the sampler tray
cover.
Continued on next page