RADIOMETER ABL800 FLEX Operator's Manual page 329

Hide thumbs Also See for ABL800 FLEX:
Table of Contents

Advertisement

ABL800 FLEX Operator's Manual
Analyzer messages,
Messages at
manager level
(continued)
Continued
No.
Message
774
Remote
operator
logged off
with user
775
Failed to
restore
Default
Setup
776
Succeeded
to restore
Default
Setup
778
Inlet gasket
should be
replaced
soon
779
Cleaning
failed
780
RADI-
ANCE
communica-
tion enabled
781
RADI-
ANCE
communica-
tion
disabled
Interpretation
An operator, remotely
logged on via NetOp,
has logged off or has
been logged off by a
local operator.
Restoring analyzer
setup to default values
has failed.
Restoring setup to
default values is
completed.
Inlet gasket is nearly
worn out and should be
replaced soon.
The last Cleaning was
aborted or interrupted.
RADIANCE
communication has
been enabled as part of
the RADIANCE
Connection Setup.
RADIANCE
communication has
been diabled as part of
the RADIANCE
Connection Setup.
11. Troubleshooting
Operator action
No action required.
Contact Radiometer service
representative.
No action required.
Replace inlet gasket.
- Check solution level and
replace container as
necessary.
- Check for and remedy
System Messages.
- Repeat Cleaning.
Removal condition:
Successful Cleaning.
Informative message. No
action required.
Informative message. No
action required.
Continued on next page