1. Introduction
Limitations of use and known interfering substances,
Continued
Carboxymethyl cellulose (CMC) Some auto-venting arterial blood samplers
CAUTION -
Known
interfering
substances
(continued)
Galactose, glucosamine, maltose,
mannose, xylose
For detailed information – see Interference Tests in chapter 5 of the ABL800 FLEX
Reference Manual .
1-6
Substance
ABL800 FLEX Operator's Manual
Interference
contain carboxymethyl cellulose (CMC) in
the porous vent. CMC can dissolve into the
sample and give erroneously low c Ca
results. Therefore we recommend
Radiometer accessories together with our
analyzers, e.g. the safe PICO arterial blood
sampler which is specifically designed to
minimize sample contamination with CMC.
Erroneously high c Glu measurements.
2+