RADIOMETER ABL800 FLEX Operator's Manual page 317

Hide thumbs Also See for ABL800 FLEX:
Table of Contents

Advertisement

ABL800 FLEX Operator's Manual
Analyzer messages,
Messages at
manager level
(continued)
Continued
No.
Message
671
Analyzing
Unit service
setup non-
default
677
Unreliable
678
Heater error One or more measuring
679
Barometer
error
Interpretation
An option in the
service setup was
changed from default
setting.
A measurement or
calibration was
attempted during
"Conditioning error in
El/Met electrode
module".
modules are over-
heated.
Measured parameter
may be unreliable due
to barometer error.
11. Troubleshooting
Operator action
Contact Radiometer service
representative.
Removal condition:
Options restored to default
values.
- Check whether
Conditioning error (525)
was reported during the last
analyzer activity. If not,
repeat measurement/
calibration.
- Check Rinse Solution
level in the container.
- Perform Rinse.
Removal condition: Rinse
without Conditioning error.
- Ensure the ambient
temperature is between 15
and 32 degrees Celsius.
- Shut down the analyzer
using the Temporary
Shutdown function, then
restart the analyzer.
- Replace fan filter if dirty.
- Shield analyzer from
direct sunlight and other
heat sources.
- Contact Radiometer
Service Representative.
Removal condition:
Analyzer temperature
within required range.
Contact Radiometer service
representative.
Continued on next page
11-35