11. Troubleshooting
Analyzer messages,
Messages at
manager level
(continued)
577-578
11-26
Continued
No.
Message
569
No Cal 1
Solution
detected by
Oxi liquid
sensor
during
initializing
571
No Rinse
Solution
detected by
OXI liquid
sensor
during
initializing
573
Oximetry
hardware
error
574
Oximetry
calibration
error
Option key
error
579
Oximetry
hardware
problem.
Not possible
to measure.
ABL800 FLEX Operator's Manual
Interpretation
No Cal 1 Solution was
detected by the Oxi
module during Startup.
No Rinse Solution was
detected by the OXI
module during Startup.
The OXI module has a
hardware error.
The measuring system
has no calibration data.
It is not possible to
make a measurement.
An error occurred in
the analyzer protection
system. The analyzer
cannot be used.
Parameter cannot be
realiably measured
because of a hardware
problem.
Operator action
Check the Cal 1 solution
for sufficient volume.
Replace as necessary.
- Perform Fluid transport
troubleshooting procedure.
- Check the Rinse solution
for sufficient volume.
Replace as necessary.
- Perform Fluid transport
troubleshooting procedure.
- Perform a 2-point
calibration.
- Contact Radiometer
Service Representative if
the problem persists.
- Perform a 2-point
calibration.
- Contact Radiometer
Service Representative.
Removal condition:
Successful 2-point
calibration.
- Shut down the analyzer
using the Temporary
Shutdown function, then
restart it.
- Contact Radiometer
Service Representative.
Removal condition:
Key installed correctly and
accepted by analyzer.
Perform a 2-point
calibration.
- Contact Radiometer
service representative.
Removal condition: No
errors detected by oximetry
hardware.
Continued on next page