11. Troubleshooting
Analyzer messages,
Messages at
manager level
(continued)
11-60
Continued
No.
Message
900
Adaptive
measuring
mode
applied
(pH/BG/
Met/El/
Crea/Oxi
[3])
901
Adaptive
measuring
mode
applied
(pH/BG/
Met/El/
Crea/Oxi
[4])
909
Met II not
filled within
time limit
919
Inlet not in
CONE3
position
before
timeout
prior to
wash with
cleaning
ABL800 FLEX Operator's Manual
Interpretation
Due to insufficient or
inhomogeneous
sample, the analyzer
has automatically
chosen an adaptive
measuring mode in
order to report as many
of the desired
parameters as possible.
Due to insufficient or
inhomogeneous
sample, the analyzer
has automatically
chosen an adaptive
measuring mode in
order to report as many
of the desired
parameters as possible.
Met II module did not
fill properly with
calibration solution.
Causes the calibration
program to abort.
Inlet was not in proper
position within
required time frame.
The activity is aborted.
Operator action
If missing parameters are
needed, repeat the
measurement with
sufficient sample free from
clots and air bubbles.
If missing parameters are
needed, repeat the
measurement with
sufficient sample free from
clots and air bubbles.
- Check the solution
containers for adequate
volume. Replace as
required.
- Perform the Fluid
Transport Troubleshooting
procedure.
Removal condition:
Successful calibration.
- Remove the inlet gasket
and inspect the condition of
inlet probe and gasket.
Replace as necessary.
Install all parts.
- Restart the analyzer.
- Contact Radiometer
Service Representative.
Removal condition:
Successful inlet calibration.
Continued on next page