Analysis setup,
Entering
The Other Fluids mode allows measurement on samples other than heparinized
corrections for
whole human blood and, therefore, requires entering corrections (offset and slope).
Other Fluids
mode
Step
Continued
Action
1.
Select the Other Fluids mode and press the Corrections button on one
of the Modes Setup screens (Capillary or Syringe).
Highlight the parameter with the arrow buttons and press Edit.
2.
3.
Key in the slope and offset values. Confirm each entry with Enter.
Please note that only values within the range programmed into the
software will be accepted.
Press Back to confirm the entries and return to the previous screen.
4.
5.
To delete a correction, highlight it and then press the
button.
3. Installation and setup
Remove
Continued on next page
3-17