ABL800 FLEX Operator's Manual
Replacing solutions and gases,
To replace a solution container, do the following:
Replacing
solutions
Step
Replacing gas
To replace a gas cylinder, do the following:
cylinders
Step
Action
1.
Remove the solution container by gripping its sides and pulling it off
the connector.
The Recording Fluid Replacements screen is displayed.
2.
Remove the cap from the new bottle.
3.
Scan the barcode on the new solution container
Always scan the barcode for each new solution container to provide
precise concentration of the installed solution.
4.
5.
Repeat steps 1-4 for the other solutions if required.
6.
For ABL8x7 FLEX analyzers only:
Press Accept to confirm that the additive has been added and to start
the proper restarting sequence or press Back to prepare the Calibration
or Cleaning Solution properly.
7.
Press Restart to start the proper restarting sequence.
Action
Press Analyzer Status – Reagents – Replace.
1.
The Recording Fluid Replacements screen is displayed.
2.
Remove the gas cylinder by turning it counterclockwise until it
becomes free.
CAUTION – Removal of safety valve
Before discarding an empty cylinder remove the safety valve
using the Valve Key (code number 922-509).
3.
Remove the cap covering the valve on the new gas cylinder. Check
that the valve is clear of any debris.
Continued
Place the new bottle in position label upward,
and push it firmly and completely onto the
connector.
7. Replacements
Continued on next page
7-17