4. Sample measurements
General information,
This mode allows you to analyze a blood sample of 35 μL and higher – up to the
FLEXMODE
maximum volume accepted by your analyzer. Depending on the available sample
volume, the FLEXMODE provides the highest number of parameters: from all
available to as many as reliably possible.
The following parameters can be obtained depending on the available sample
volume (for the ABL835):
35-40 μL*
40-55 μL*
50-70 μL*
65-100 μL*
> 90 μL*
195 μL
* All volumes are approximate volumes at ctHb = 15 g/dL.
** The messages will accompany the obtained result – see Patient result messages
at the end of this chapter. For interpretation of the messages – see Troubleshooting
messages in chapter 11 in this manual.
1.
CAUTION
notices
2.
3.
4.
4-6
Continued
Volume
pH + OXI
pH + OXI
pH + OXI + BG
pH + OXI + BG + MET
pH + OXI + BG + MET + EL
pH + OXI + BG + MET + EL + ctBil
CAUTION – Presence of glucolic acid
Never use analyzer for lactate measurements if there is any
suspicion of presence of glycolic acid in blood as e.g. in case of
ethylene glycol poisoning or xylitol infusion. Glycolic acid
interferes with the lactate sensor resulting in erroneously high
lactate readings
CAUTION – Risk of incorrect result
Visually inspect the measuring chambers for blockages when lifting
the inlet flap prior to performing a measurement. If there is a
blockage, close the flap and the analyzer will perform a rinse. If
repeated visual inspection shows blockages in the measuring
chamber, clean the electrodes and the measuring chamber according
to step 7 in chapter 11: Troubleshooting, section "Fluid transport
troubleshooting procedure" in this manual.
CAUTION - Risk of incorrect result
Always remove air bubbles from a sample before introducing it into
the analyzer. Air bubbles may result in erroneous pO
CAUTION - Risk of incorrect result
During measurement, use the sample path view to observe the
sample in the measuring chambers. If any bubbles are seen,
disregard the result and repeat the measurement.
ABL800 FLEX Operator's Manual
Parameter profiles
Message No. **
869
870
871, 872
873
874
No message
values.
2
Continued on next page