ABL800 FLEX Operator's Manual
tHb calibration,
tHb calibration
To verify the validity of the tHb calibration, do the following.
check
Step
* Can be ctHb, or ctHb and
NOTICE:
If the tHb calibration check has not been performed immediately after the tHb
calibration, repeat the calibration and the calibration check immediately after it.
Continued
Action
1.
Open the syringe inlet flap and place the adapter with the S7770
solution in the inlet.
Select the mode: Syringe – S195 μL or Syringe – S85 μL or
2.
Syringe – S250 µL depending on the ABL800 FLEX version.
3.
Press the Start button. Remove the ampoule when prompted and close
the inlet.
4a.
If the calibration check value* is within the insert limits, the tHb
calibration is accepted.
4b.
If the calibration check value* is outside the insert limits, repeat the
tHb calibration and then the calibration check immediately after it.
If the second tHb calibration and the calibration check fail again,
contact a service technician.
ctBil,
or ctBil.
6. Calibration
6-7