Measured
New table for pleural liquid:
parameters
* If the measured values obtained lie outside the test range,
Radiometer advises you to repeat the measurement by means of
another method.
Environmental
Indoor use statement added according to CSA regulations:
requirements
Location
Environmental
Pollution degree 2 statement added according to CSA regulations:
ratings
Installation category II.
Pollution degree 2.
Clot detection
A new feature is included in software version 6.06 of the following
products:
The new feature enables the analyzer to detect clots caught in the
measurement chamber under the pH electrode. This document
describes it.
When to activate the clot-detection feature
It is especially beneficial to activate the clot detection feature in
analyzers frequently used to analyze samples known to be prone to
clotting, for example, samples drawn from umbilical cords. Once
activated clot detection is active during all sample measurements.
NOTICE: It is important to note that activating the feature will
delay measurement results by approximately one minute, even
though it only increases the measurement cycle time by five
seconds. It also increases consumption of Rinse solution, but this is
minimal.
Parameter
Unit
pH
pH scale
 ABL7xx Series (XPE) analyzers
 ABL800 FLEX analyzers
 ABL800 BASIC analyzer.
Measuring
Range
6.300-8.000
Indoor use only
Test
range
7.0-7.5*