RADIOMETER ABL800 FLEX Operator's Manual page 425

Hide thumbs Also See for ABL800 FLEX:
Table of Contents

Advertisement

ABL800 FLEX Operator's Manual
Calibration verification,
Preparatory
Step
steps (continued)
Verification
To perform a verification measurement, do the following:
procedure
Step
Once the first measurement in this mode has been performed, the mode is available
for the next seven days. During this time you can verify your reportable ranges
according to the procedure used in your institution. After seven days the mode will
not be available for the next 14 days.
Continued
Action
5.
Condition calibration verification kit – see Preparing a control
solution in chapter 5 of this manual.
Place the ampoule in the ampoule opener and break off the ampoule
6.
neck. Place the ampoule fully into the H700 Adapter.
Action
1.
Check that the analyzer is in the Ready mode.
2.
Open the syringe inlet flap and place the adapter tip into the syringe
inlet.
3.
Select the Cal. verification mode on the screen and press Start.
Press Aspirate to start sample aspiration.
4.
When prompted by the analyzer, remove the sampling device and close
5.
the inlet flap.
Enter the information on the Patient Identification screen.
6.
Use a dedicated identifier for each solution (e.g. Level 1) as Patient ID.
After all measurements on the calibration verification solutions have been
7.
completed, review the data by filtering it in the Patient Results log or
transfer the data to a data analysis program – see Exporting Data Logs in
chapter 8: Disk Functions.
After the new reportable ranges have been determined, enter any changes
8.
in Setup program: Reportable Ranges.
15. Radiometer settings
15-19