ABL800 FLEX Operator's Manual
Quality control setup,
Rilibäk Ranges
Step
(continued)
To edit a Rilibäk rule, do the following:
Step
To remove a Rilibäk rule, do the following:
Step
NOTICE: When a quality control measurement violates an applied Rilibäk Rule, a
red R is shown in front of the parameter name in the result.
Continued
Action
7.
To select the desired +/- range press the desired radio button.
8.
Enter the desired +/- range in the "Ranges" box.
9.
Press Back to return to the Rilibäk Ranges screen. The added Rilibäk
rule is now shown in the screen.
Action
1.
Select the desired rule in the Rilibäk Ranges screen and press Edit to
display the screen below:
2.
Use the Right or Left arrows to jump between the input boxes and edit
the desired values.
3.
Press Back to return to the Rilibäk Ranges screen.
Action
1.
Highlight the desired rule in the Rilibäk Ranges screen and press
Delete.
3. Installation and setup
3-47