RADIOMETER ABL800 FLEX Operator's Manual page 89

Hide thumbs Also See for ABL800 FLEX:
Table of Contents

Advertisement

ABL800 FLEX Operator's Manual
Analysis setup,
Sample Pre-
Step
registration
setup
(continued)
(cont.)
This program allows you to select the following: the batch mode (automatic
Sample Logistics
Setup
sample handling), Patient Identification shown on the Ready screen and to define
maximum allowed sample age for each measured parameter.
Batch mode
The batch mode allows you to place your sampler in an empty slot in the FLEXQ
(automatic
sampler tray and leave the analyzer. The sampler will be registered, sample
sample
analyzed and the results sent to the required person.
handling)
To achieve this the following is required:
• Analyzer has live connection to the RADIANCE system
• Barcode interpretation is set to Sampler ID in the Sample Pre-registration
• "When entering Sampler ID" in "Request patient demographics" box is selected
and "From connection" must be set to RADIANCE in the Automatic Data
Request setup
• All sample/patient information is present on the RADIANCE system
Continued
Action
Note that choosing the Accession Number or Sampler ID will gray out
1
its check button (Sampler ID on the screen above). Choosing the
Accession Number or Sampler ID ensures that this data will be
included in the "Data confirmation" box.
2.
Select the barcode entry.
3.
Activate the relevant check buttons in the "Included fields" box:
Accession Number, Sampler ID, Patient First Name, Patient Last
Name, Birth Date, Patient Sex.
4.
Activate the check button in the "Data confirmation" box so that the
patient data can be confirmed before the sample measurement.
Press Close to confirm the settings and return to the Ready screen.
5.
3. Installation and setup
Continued on next page
3-23