RADIOMETER ABL800 FLEX Operator's Manual page 351

Hide thumbs Also See for ABL800 FLEX:
Table of Contents

Advertisement

ABL800 FLEX Operator's Manual
Analyzer messages,
Messages at
manager level
(continued)
963, 964
Continued
No.
Message
961
Total
concentration
outside
measuring
range
962
Ambient
temperature
not defined
Leak
currents
detected
965
Cleaning –
Additive is
expired
Interpretation
The total concentration
of Crea + Creatine is
above the measuring
range and the analyzer
is not able to measure
Crea within the
specifications.
See chapter 14:
Specifications in the
Operator's Manual.
Ambient temperature
has not been defined by
the user in
Environment Setup.
Until the ambient
temperature has been
defined, a default value
of 25 °C is used during
High Crea Check and
the internal Crea
calibration calculations.
Please maintain an
ambient temperature
between 18 °C and 29
°C
Leak currents are
detected during
calibration and may
distort the measuring
results. The affected
parameters are marked
with "?".
The Cleaning additive
is too old and the
Cleaning Solution
cannot be used
anymore.
11. Troubleshooting
Operator action
- Check for and remedy
other errors related to the
result, system messages or
calibration status.
- Perform a High Crea
Check. If the High Crea
Check result is accepted,
the blood sample may be
suspected.
- Perform measurement on
new blood sample.
- Perform the Electrode
Troubleshooting procedure.
Enter ambient temperature
in Environment Setup - see
Operator's Manual chapter
3.
Removal condition:
Ambient temperature
entered.
Contact Radiometer service
representative.
Replace Cleaning Solution
bottle – remember to add
the additive.
Continued on next page
11-69