RADIOMETER ABL800 FLEX Operator's Manual page 311

Hide thumbs Also See for ABL800 FLEX:
Table of Contents

Advertisement

ABL800 FLEX Operator's Manual
Analyzer messages,
Messages at
manager level
(continued)
Continued
No.
Message
605
Cal expired
606
Cal expired
(pH)
608
Cal expired
(pCO
)
2
609
Cal expired
(pO
)
2
610
Cal expired
(K)
611
Cal expired
(Na)
612
Cal expired
(Ca)
613
Cal expired
(Cl)
614
Cal expired
(Glu)
Cal expired
615
(Lac)
616
Cal expired
(OXI)
618
Waste
detector
error
619
Waste
container
missing
Interpretation
Too long time passed
since the last successful
calibration of the
parameter. Calibrations
of the parameter are
considered invalid and
values reported as
".....".
The waste detector may
be faulty or damaged.
The waste container is
missing or improperly
mounted.
11. Troubleshooting
Operator action
Perform 1- or 2-point
calibration.
Removal condition:
Successful 1- or 2-point
calibration.
Contact Radiometer service
representative.
Removal condition:
Functioning waste detector
installed.
Mount the waste container
properly.
Removal condition: Waste
container properly
mounted.
Continued on next page
11-29