11. Troubleshooting
Analyzer messages,
Messages at
manager level
(continued)
529, 531-
532, 534-
11-24
Continued
No.
Message
512
Temperatu-
re error
513
No gas
521
Inhomoge-
neous
sample
Liquid
sensor
537
calibration
error
ABL800 FLEX Operator's Manual
Interpretation
The temperature was
outside the required
range during
measurement or
calibration.
All results are marked
with "?".
Measurement or
calibration was
performed with Gas
error present.
Activity is aborted.
Air bubbles were
detected in the sample.
Results may have "?".
One or more of the
liquid sensors failed to
calibrate.
Operator action
Ensure ambient
temperature is between 15
and 32 °C.
- If the analyzer has
recently performed a Cold
start, wait for temperature
error to disappear.
- Shield analyzer from
direct sunlight or heat
sources.
- Contact Radiometer
service representative.
- Check preasure in gas
cylinders. Replace as
required.
- Verify that cylinders are
properly installed.
- Enter the Replacement
screen (via Replacement
Status) and re-scan the
barcode labels of the gas
cylinders.
- Ensure sampler and
sample volume match the
selected sample mode.
- Repeat measurement.
- Check pump tubes and
inlet for leaks. Replace
tubing and inlet gasket as
required.
- Repeat liquid sensor
calibration.
- Perform Fluid transport
troubleshooting procedure.
- Contact Radiometer
Service Representative.
Continued on next page