RADIOMETER ABL800 FLEX Operator's Manual page 331

Hide thumbs Also See for ABL800 FLEX:
Table of Contents

Advertisement

ABL800 FLEX Operator's Manual
Analyzer messages,
Messages at
manager level
(continued)
Continued
No.
Message
790
Adaptive
measuring
mode
applied
791
Parameter
could not be
measured
792
Liquid
sensor
calibration
error
793
pH/BG
pump
asymmetry
error
794
El/Met
pump
asymmetry
error
Interpretation
Due to insufficient or
inhomogeneous
sample, the analyzer
has automatically
chosen an adaptive
sample handling in
order to report as many
of the desired
parameters as possible.
Due to insufficient or
inhomogenous sample,
the parameter in
question could not be
measured.
Inlet lower liquid
sensor failed to
calibrate.
The pH/BG module
pump failed to calibrate
pump asymmetry. The
same calibration values
will be used for both
directions until a
successful calibration
of pump asymmetry is
performed.
The El/Met module
pump failed to calibrate
pump asymmetry. The
same calibration values
will be used for both
directions until a
successful calibration
of pump asymmetry is
performed.
11. Troubleshooting
Operator action
If missing parameters are
needed, repeat the
measurement with
sufficient sample free of
clots and air bubbles.
If missing parameters are
needed, repeat the
measurement with
sufficient sample free of
clots and air bubbles.
- Repeat liquid sensor
calibration.
- Perform the Fluid
Transport Troubleshooting
procedure.
- Contact Radiometer
service representative.
- Perform the Pump
troubleshooting procedure.
- Contact Radiometer
service representative.
- Perform the Pump
troubleshooting procedure.
- Contact Radiometer
service representative.
Continued on next page
11-49