11. Troubleshooting
Analyzer messages,
Messages at
manager level
(continued)
11-50
Continued
No.
Message
795
Met II pump
asymmetry
error
796
Solution
pump
asymmetry
error
797
Waste pump
asymmetry
error
798
User logged
on
799
User logged
off
800
Logon
attempt
failed
801
Hard disk
S.M.A.R.T.
threshold(s)
exceeded
ABL800 FLEX Operator's Manual
Interpretation
The Met II module
pump failed to calibrate
pump asymmetry. The
same calibration values
will be used for both
directions until a
successful calibration
of pump asymmetry
has been performed.
The Solution module
pump failed to calibrate
pump asymmetry. The
same calibration values
will be used for both
directions until a
successful calibration
of pump asymmetry is
performed.
The Waste pump in the
Solution module failed
to calibrate pump
asymmetry. The same
calibration values will
be used for both
directions until a
successful calibration
of pump asymmetry is
performed.
User logged on
successfully.
User logged off.
User tried to log on but
did not provide a valid
password.
The analyzer's hard
disk drive has given a
warning. This may
mean imminent hard
disk drive failure.
Operator action
- Perform the Pump
troubleshooting procedure.
- Contact Radiometer
service representative.
- Perform the Pump
Troubleshooting procedure.
- Contact Radiometer
service representative.
- Perform the Pump
Troubleshooting procedure.
- Contact Radiometer
service representative.
No action required.
Informative message.
Informative message. No
action required.
Provide a valid password to
log on.
Contact Radiometer service
representative.
Continued on next page