RADIOMETER ABL800 FLEX Operator's Manual page 348

Hide thumbs Also See for ABL800 FLEX:
Table of Contents

Advertisement

11. Troubleshooting
Analyzer messages,
Messages at
manager level
(continued)
11-66
Continued
No.
Message
951
Creatine is
below the
reportable
range
952
Creatine is
above the
reportable
range
953
Measured
QC value
above the
control
range
(Crea A)
954
Measured
QC value
below the
control
range
(Crea A)
ABL800 FLEX Operator's Manual
Interpretation
Creatine is below the
reportable range.
See chapter 14:
Specifications in the
Operator's Manual.
Creatine is above the
reportable range.
See chapter 14:
Specifications in the
Operator's Manual.
Creatine value is above
the control range.
Error on the Creatine
measurement is
possible.
Creatine value is below
the control range.
Error on the Creatine
measurement is
possible.
Operator action
- Check for and remedy
other errors related to the
result, system messages or
calibration status.
- Perform QC. If the QC
result is accepted, the
blood sample may be
suspected.
- Perform measurement on
new blood sample.
- Perform the Electrode
Troubleshooting procedure.
- Check for and remedy
other errors related to the
result, system messages or
calibration status.
- Perform QC. If the QC
result is accepted, the
blood sample may be
suspected.
- Perform measurement on
new blood sample.
- Perform the Electrode
Troubleshooting procedure.
- Verify procedure and
repeat measurement.
- Refer to Quality Control
Systems Reference
Manual.
- Verify procedure and
repeat measurement.
- Refer to the Quality
Control Systems Reference
Manual.
Continued on next page