ABL800 FLEX Operator's Manual
Analyzer messages,
Messages at
manager level
(continued)
Continued
No.
Message
941
Cal expired
(Crea)
942
Cal expired
(Creatine)
944
Met II
module not
expected
945
Met II
module not
active
946
Conditioning
error in Met
II electrode
module
Interpretation
Too long time passed
since the last successful
calibration of the
parameter. Parameter
measurement values
are reported as ".....".
Software configuration
does not match
analyzer type.
The Met II module is
not responding due to
internal communication
problems, or software
configuration does not
match the analyzer
type.
Rinse Solution was not
detected in Met II
electrode module. The
Crea results will be
reported as "....." in the
next measurement.
11. Troubleshooting
Operator action
Perform a 2-point
calibration.
Removal condition:
Successful 2-point
calibration.
Contact Radiometer
Service Representative.
Removal condition:
Matching software and
hardware configuration.
- Shut down the analyzer
using the Temporary
Shutdown function, then
restart it.
- Contact Radiometer
Service Representative.
Removal condition:
Met II module ready, or
software configured
without Met II module
support.
- Check Rinse Solution for
adequate volume, proper
installation or obstructions
in bottle opening.
- Replace as necessary.
- Perform the Fluid
Transport Troubleshooting
procedure.
Removal condition:
Succesful Rinse.
Continued on next page