7. Replacements
Replacing solutions and gases,
Replacing gas
Step
cylinders
(continued)
7-18
Action
4.
Scan the barcode on a new gas cylinder.
5.
6.
The replaced items appear in the "Recorded replacements" box as
soon as the code has been scanned or typed.
To remove an item from the recorded replacements list, highlight it
with the up/down arrow buttons in the box and press Undo.
7.
Press Restart to start the proper restarting sequence.
CAUTION – Risk of personal injury
Pressurized container. Non-flammable compressed gas. Do not breathe
gas. Gas mixtures containing less than 19.5 % oxygen may cause
suffocation. Protect from sunlight and do not expose to temperatures
exceeding 50 °C. Store and use with adequate ventilation. Keep away from
oil and grease. Do not refill.
ABL800 FLEX Operator's Manual
Continued
Place the gas cylinder valve into position on
the socket of the regulator. To ease alignment,
use the analyzer enclosure as a guide.