3. Installation and setup
Quality control setup,
Westgard Rules
Step
(continued)
3-44
Continued
Action
Press Next Slot and activate/deactivate the Westgard Rules for the
2.
other slots in the same manner.
3.
To activate the Westgard rules for specific parameters in a slot,
display the desired slots by using Next Slot.
4.
Press the Edit button. The screen shows all the Westgard Rules and
their current settings for the parameter.
5.
Display the desired parameter using the Next Param or Prev Param
buttons.
6.
Activate the desired Westgard Rule(s) by pressing the corresponding
check button.
Once edited, all future quality control measurements on the parameter
for the given slot will be evaluated according to the selected Westgard
Rule(s).
7.
Select Westgard Rules for other parameters or levels in the same
manner.
8.
To select/deselect all six Westgard Rules for every parameter of a
displayed control solution slot, select the desired slot by pressing the
Next Slot button.
ABL800 FLEX Operator's Manual
Parameter
Continued on next page