11. Troubleshooting
Forced Hold causes,
Inlet calibration
Step
error
(continued)
Leak Detected
This condition occurs when a leakage test has failed, indicating a leak in the fluid
error
transport system.
To correct the error, do the following:
Step
11-6
Continued
Action
4.
Check the condition of the inlet probe. If it appears bent or damaged,
replace it.
5.
Mount all components, then restart the analyzer by pressing Restart.
If the error has been corrected, the analyzer performs an inlet
calibration during restart and enters the Ready mode.
If the Hold mode remains, contact a service representative.
Action
1.
Remove the analyzer covers and inspect the fluid transport system for
leaks.
2.
Check for poor tube connections or worn tubes. Repair as required.
3.
Check the inlet gasket; change if required.
4.
Ensure that electrodes are properly mounted in the measuring chamber
and the electrode connector is latched shut.
5.
Mount all items, restart the analyzer by pressing Restart.
If the error has been corrected, the analyzer performs a restart
sequence (includes a leakage test) and enters the Ready mode.
If the Hold mode remains, contact a service representative.
ABL800 FLEX Operator's Manual