11. Troubleshooting
Analyzer messages,
Messages at
manager level
(continued)
11-62
Continued
No.
Message
926
Inlet's liquid
sensor to
Oxi's liquid
sensor
volume
could not be
confirmed 3
927
Inlet's liquid
sensor to
Oxi's liquid
sensor
volume
could not be
confirmed 4
928
Unable to
fill Inlet
with
continuous
rinse during
pump
calibration
929
TIPCAP
conflict
detected 1
930
TIPCAP
conflict
detected 2
ABL800 FLEX Operator's Manual
Interpretation
During pump
calibration,
measurement of
internal volume part
failed.
Causes the Fluid
Transport Control
program to abort.
The detection of the
TIPCAP failed.
The activity is aborted.
Operator action
- Check Rinse Solution for
sufficient volume. Replace
as necessary.
- Perform Pump
Calibration.
- Perform the Fluid
Transport Troubleshooting
procedure.
Removal condition:
Successful pump
calibration.
- Remove the inlet flaps
and the inlet gasket.
Inspect the condition of the
inlet probe and gasket.
Look for obstructions that
may hinder probe
movement. Replace as
necessary. Install all parts.
- Perform the Inlet Probe
Troubleshooting
Procedure.
- Restart the analyzer.
- Contact Radiometer
Service Representative.
Removal condition:
Successful inlet calibration.
Continued on next page