RADIOMETER ABL800 FLEX Operator's Manual page 165

Hide thumbs Also See for ABL800 FLEX:
Table of Contents

Advertisement

ABL800 FLEX Operator's Manual
Measurements with FLEXQ,
An urgent
To bypass a queued measurement while the analyzer processes the samplers in the
measurement
sampler tray, do the following:
Step
Note that if the Urgent Manual Sample button is grayed out, the inlets are free and
the sample can be analyzed immediately.
To change or add information to sample registration, do the following:
Editing sample
registration
Step
Action
1.
Check that Urgent Manual Sample is active on the Ready screen and
press it.
2.
If the urgent measurement violates parameter lifetime of one or more
queued samples, accept violation and book the inlet for manual
measurement by pressing Yes or abort booking of the inlet by pressing
No.
3.
When allowed on the screen, open the inlet flap, insert the sampling
device in the inlet and press Start – see Introducing a blood sample in
this chapter.
Action
Locate the required sample in the "FLEXQ Status" on the Ready
1.
screen and press the
Registration screen.
Change the patient data on the Sample Registration screen.
2.
3.
Press Accept to confirm the patient data.
Continued
button to proceed to the Sample
4. Sample measurements
4-11