RADIOMETER ABL800 FLEX Operator's Manual page 309

Hide thumbs Also See for ABL800 FLEX:
Table of Contents

Advertisement

ABL800 FLEX Operator's Manual
Analyzer messages,
Messages at
manager level
(continued)
Continued
No.
Message
580
Oximetry
hardware
problem.
Possible to
measure.
581
Oximetry
measuring
error
582
tHb
calibration
outside
limits
583
Measured
value
outside
reportable
range for
the
parameter
588
Measured
QC value
lower than
statistical
range
Interpretation
A hardware problem
exists in the OXI
module. Parameter can
be measured.
Spectrum deviates from
expected blood or QC
spectrum.
Measurement may be
unrealiable.
tHb calibration failed.
The parameter value is
outside the defined
reportable range or the
analyzer measurement
range.
The parameter value is
below the lower limit
of the user-defined
statistical range.
Measurement is not
included in statistics.
11. Troubleshooting
Operator action
- Perform a 2-point
calibration.
- Contact Radiometer
service representative.
Removal condition: No
errors detected by oximetry
hardware.
- Check patient record for
medication containing
possible interfering
substances.
- Call a 2-point calibration.
- Contact Radiometer
Service Representative.
Repeat tHb calibration.
Removal condition:
Successful tHb calibration.
- Check for and remedy
other errors in the result,
and any errors in System
Messages or Calibration
Status.
- Perform QC. If the QC
result value lies within the
measuring range, the blood
sample may be suspected.
- Perform the Electrode
Troubleshooting procedure.
- Verify procedure and
repeat the measurement.
- Refer to the Quality
Control Systems Reference
Manual for details on the
evaluation of the results.
Continued on next page
11-27