Corrective actions,
Explanation of
Corrective action
corrective
actions
Re-run same level
(continued)
twice
Use ampoule last
in schedule
NOTICES:
• The critical system messages will always result in a red traffic light signal.
• If the analyzer locks due to QC or replacement schedule reminders, a critical
Continued
If an error has been registered during a scheduled
AutoCheck measurement, it will be repeated twice
provided the control solution is available in the carousel.
If a control solution scheduled for AutoCheck is missing in
the carousel, the measurement will be on the control
solution type scheduled before this one.
system message will be generated and the traffic light signal will be red
The specified traffic light signal and the messages will continue to appear
until the condition no longer exists.