Analyzer messages,
Messages at
manager level
(continued)
Continued
No.
Message
654
Sample in
El/Met
upper liquid
sensor
detected
prematurely
during
aspiration
655
Sample in
pH/BG
upper liquid
sensor
detected
prematurely
during
aspiration
656
Sample in
OXI liquid
sensor
detected
prematurely
during
aspiration
658
Inlet flaps
removed
659
Cover
removed or
window
opened
660
Inlet
calibration
error
Interpretation
El/Met module was not
filled correctly during
sample aspiration.
pH/BG module was not
filled correctly during
sample aspiration.
OXI module was not
filled correctly during
sample aspriration.
The analyzer entered
HOLD because both
inlet flaps were
removed.
The analyzer entered
HOLD because the
cover was removed or
the window to the
electrode modules
opened.
An error occurred
during Inlet calibration.
Analyzer entered
HOLD mode.
11. Troubleshooting
Operator action
- Ensure sample is free of
clots.
- Repeat measurement.
- Perform Fluid transport
troubleshooting procedure.
-Ensure sample is free of
clots.
- Repeat measurement.
- Perform Fluid transport
troubleshooting procedure.
- Ensure sample is free of
clots.
- Repeat measurement.
- Perform the Fluid
Transport Troubleshooting
procedure.
Ensure the inlet flaps are
properly mounted prior to
restart.
Removal condition: Inlet
flaps properly mounted.
Position the cover and
close the window prior to
restart.
Removal condition: Cover
in position and window
closed.
- Remove the inlet gasket. -
Inspect the gasket and inlet
probe. Replace as
necessary.
- Press Restart.
Removal condition:
Successful Inlet calibration
Continued on next page