RADIOMETER ABL800 FLEX Operator's Manual page 325

Hide thumbs Also See for ABL800 FLEX:
Table of Contents

Advertisement

ABL800 FLEX Operator's Manual
Analyzer messages,
Messages at
manager level
(continued)
Continued
No.
Message
DMS not
734
connected to
Analyzing
Unit
738
Ampoule(s)
in
AutoCheck
carousel not
installed
739
Solution's
composition
not
identifiable
744
AutoCheck
Laser Setup
Error
745
Low hard
disk space
746
Rinse
Solution
low warning
Interpretation
Data management
system establishes
connection to the
Analyzing Unit, or the
connection is lost.
AutoCheck carousel
contains an ampoule
which was not selected
in Control Solutions
Setup program.
AutoCheck ampoule
solutions do not match
labeling.
AutoCheck barcode
reader has failed.
Free hard disk space is
below 100 Mb.
Rinse Solution level is
below the defined
warning level.
11. Troubleshooting
Operator action
- Wait a few minutes for
the connection to establish.
- Restart the analyzer.
- If the error persists,
contact Radiometer service
representative.
Read in the barcodes for all
ampoules present in
AutoCheck carousel – use
inserts.
Removal condition:
Barcodes read in for all
ampoules present in the
AutoCheck carousel.
- Repeat measurement.
- Contact Radiometer
service representative if the
problem persists.
- Restart analyzer.
- If the error persists,
contact Radiometer service
representative.
Move archive files to
another storage device.
Removal condition: More
than 100 Mb of free hard
disk space.
Replace container as
required.
Removal condition:
Solution level above
warning level.
Continued on next page
11-43