11. Troubleshooting
Analyzer messages,
Messages at
manager level
(continued)
11-10
Continued
No.
Message
134
LIS/HIS:
Connection
established
165
LIS/HIS:
High level
protocol
could not
generate
high level
packet
166
LIS/HIS:
General
communica-
tion error
167
LIS/HIS:
High level
protocol
received
packet in
wrong
format
200
User msg:
201
Westgard
Rule (1.2s)
violation
202
Westgard
Rule (1.3s)
violation
203
Westgard
Rule (2.2s)
violation
ABL800 FLEX Operator's Manual
Interpretation
The connection was
successfully
established.
An error occurred
while formatting a
message.
An internal error
occurred in the
LIS/HIS
communication
module.
An error occurred
while parsing
(interpreting) a
message.
This is only a message.
An operator has
entered a note in the
log.
Measured parameter
value is outside the
mean ± 2 SD range.
Measured parameter
value is outside the
mean ± 3 SD range.
Two consecutive
measurements are
outside the mean ± 2
SD range on the same
side of the mean. This
may indicate a shift.
Operator action
No action required. The
message is informative.
- Check protocol
configurations.
- Contact Radiometer
service representative.
Contact Radiometer service
representative if the
problem persists.
- Check protocol
configurations.
- Contact Radiometer
service representative.
No action required.
- Verify procedure and
repeat measurement.
- Check Replacement
Status for pending
replacements (including
elctrodes).
- Refer to the Quality
Control Systems Reference
Manual for detailed
evaluation procedure.