No.
Message
218
Inlet
positioning
error
219
Condition-
ing error in
El/Met
electrode
module
232
Oximetry
calibration
error
Oximetry
hardware
error
316
Temperature
error
Interpretation
Inlet probe was not
positioned properly
within required time
frame.
Causes the activity to
abort.
Rinse Solution was not
detected in El/Met
electrode module. The
glucose and lactate
results will be reported
as "....." in the next
measurement.
Water spectrum
intensity was too low
or too high during
absorbance calculation.
OXI measurements
cannot be performed.
The Oxi module has a
hardware error.
Analyzer temperature
is outside 37.0 ± 0.2
degrees Celsius.
11. Troubleshooting
Operator action
Perform the Inlet
Troubleshooting Procedure
Removal condition:
Successful inlet calibration.
- Check Rinse Solution for
adequate volume, proper
installation or obstructions
in container opening.
Replace as necessary.
- Perform the Fluid
transport troubleshooting
procedure.
Removal condition:
Successful Rinse.
- Perform a 2-point
calibration.
- Contact Radiometer
Service Representative if
the problem persists.
Removal condition:
Successful 2-point
calibration.
- Perform a 2-point
calibration.
- Contact Radiometer
Service Representative if
the problem persists.
- Ensure the ambient
temperature is between 15
and 32 degrees Celsius.
- If the system has just
performed a cold start, wait
for error to disappear.
- Replace the fan filter if
dirty.
- Shield analyzer from
direct sunlight and other
heat sources.
- Contact Radiometer
service representative.