11. Troubleshooting
Analyzer messages,
Messages at
manager level
(continued)
11-54
Continued
No.
Message
869
Adaptive
sample
handling
applied
(decay to
pH/Oxi
without Bil)
870
Adaptive
sample
handling
applied
(decay to
pH/Oxi)
871
Adaptive
sample
handling
applied
(decay to
pH/BG/Oxi)
872
Adaptive
measuring
mode
applied
(pH/BG/Oxi
[2])
873
Adaptive
measuring
mode
applied
(pH/BG/
Met/Oxi)
874
Adaptive
measuring
mode
applied
(decay to all
micro
mode)
875
Sample
aged
ABL800 FLEX Operator's Manual
Interpretation
Due to insufficient or
inhomogeneous
sample, the analyzer
has automatically
chosen an adaptive
measuring mode in
order to report as many
of the desired
parameters as possible.
The specified limit for
sample age has been
exceeded.
Operator action
If missing parameters are
needed, repeat the
measurement with
sufficient sample free from
clots and air bubbles.
Draw and analyze new
sample.
Continued on next page