Analyzer messages,
Messages at
manager level
(continued)
Continued
No.
Message
538
pH/BG
pump error
539
El/Met
pump error
540
Met II pump
error
541
Solution
pump error
564
No
Cleaning
Solution
detected by
Oxi liquid
sensor
during
initializing.
567
No Cal 2
Solution
detected by
Oxi liquid
sensor
during
initializing
Interpretation
The pH/BG module
pump failed to
calibrate. Previous
calibration is used until
a successful pump
calibration is
performed.
The El/Met module
pump failed to calibra-
te. Previous calibration
is used until a success-
ful pump calibration is
performed.
The Met II module
pump failed to
calibrate. Previous
calibration is used until
a successful pump
calibration is
performed.
The solution pump
failed to calibrate.
Previous calibration is
used until a successful
pump calibration is
performed.
No Cleaning Solution
was detected by Oxi
module during Startup.
No Cal 2 Solution was
detected by Oxi module
during Startup.
11. Troubleshooting
Operator action
- Perform Pump
troubleshooting procedure.
- Contact Radiometer
service representative.
- Perform Pump
troubleshooting procedure.
- Contact Radiometer
service representative.
- Perform Pump
troubleshooting procedure.
- Contact Radiometer
service representative.
- Perform Pump
Troubleshooting procedure.
- Contact Radiometer
service representative.
- Check the Cleaning
solution for sufficient
volume. Replace as
necessary.
- Perform Fluid transport
troubleshooting procedure.
- Check the Cal 2 solution
for sufficient volume.
Replace as necessary.
- Perform Fluid transport
troubleshooting procedure.
Continued on next page
11-25