11. Troubleshooting
Analyzer messages,
Messages at
manager level
(continued)
11-30
Continued
No.
Message
620
Waste
container
full
621
Waste
container
critically
full
Expected
622
liquid level
below 0
623
Solution
empty
624
Solution
missing
ABL800 FLEX Operator's Manual
Interpretation
Waste container is
filled to near-maximum
capacity. Unless
replaced soon, the
analyzer will enter
Standby until the
container is replaced.
The waste container is
filled to maximum
capacity. The analyzer
enters Standby mode;
no measurements or
calibrations can be
performed.
One or more of the
solution barcodes was
entered incorrectly
during replacement.
The analyzer cannot
monitor solution
consumption.
Wet section activities
are suspended due to
one or more empty
solution containers.
One or more solution
containers is missing or
improperly mounted.
Operator action
Replace waste container.
Removal condition: Empty
Waste container installed.
Replace waste container.
Removal condition: Empty
Waste container installed.
Check the solution
containers and replace as
necessary. Remember to
scan barcodes.
Removal condition:
Solution containers
correctly installed and
solution barcodes correctly
recorded by the analyzer.
Check the solution levels
and replace containers as
necessary.
Removal condition:
Sufficient level of all
solutions.
Check that all solution
containers are mounted
properly.
Removal condition: All
solution containers are
mounted correctly.
Continued on next page