ABL800 FLEX Operator's Manual
Analyzer messages,
Messages at
manager level
(continued)
Continued
No.
Message
894
Adaptive
measuring
mode
applied
(pH/BG/
Met/Oxi
[2])
895
Adaptive
measuring
mode
applied
(pH/BG/
Met/El/Oxi)
896
Adaptive
measuring
mode
applied
(pH/BG/
Met/El/Oxi
[2])
897
Adaptive
measuring
mode
applied
(pH/BG/
Met/El/Oxi
[3])
898
Adaptive
measuring
mode
applied
(pH/BG/
Met/El/Crea
/Oxi)
899
Adaptive
measuring
mode
applied
(pH/BG/
Met/El/
Crea/Oxi
[2])
Interpretation
Due to insufficient or
inhomogeneous
sample, the analyzer
has automatically
chosen an adaptive
measuring mode in
order to report as many
of the desired
parameters as possible.
11. Troubleshooting
Operator action
If missing parameters are
needed, repeat the
measurement with
sufficient sample free from
clots and air bubbles.
Continued on next page
11-59