RADIOMETER ABL800 FLEX Operator's Manual page 310

Hide thumbs Also See for ABL800 FLEX:
Table of Contents

Advertisement

11. Troubleshooting
Analyzer messages,
Messages at
manager level
(continued)
11-28
Continued
No.
Message
589
Measured
QC value
higher than
statistical
range
593
Insufficient
sample
597
Liquid
detected in
pH/BG
module
during Gas
Calibration
Demo
600
option will
expire soon
604
Parameter
not installed
ABL800 FLEX Operator's Manual
Interpretation
The parameter value is
above the upper limit
of the user-defined
statistical range.
Measurement not
included into statistics.
Sample volume is too
small for the selected
measuring mode.
Affected parameters
will be marked with
"?".
Unexpected liquid is
detected by the pH/BG
module during
calibration. Calibration
is aborted.
One or more currently
installed demo options
will expire and be
removed within the
next 7 days.
Parameter was not
installed or is
corrupted. Parameter
cannot be measured.
Operator action
- Verify procedure and
repeat the measurement.
- Refer to the Quality
Control Systems Reference
Manual for details on the
evaluation of the results.
- Make sure sample volume
matches the selected
measuring mode.
- Repeat the measurement,
ensuring sufficient sample
volume.
- Contact Radiometer
service representative.
- Repeat calibration.
- Perform Fluid transport
troubleshooting procedure.
To have the option
permanently, contact a
Radiometer sales or service
representative.
Removal condition:
Options expired or
permanent options
installed.
Contact Radiometer service
representative.
Continued on next page