11. Troubleshooting
Analyzer messages,
Messages at
manager level
(continued)
11-44
Continued
No.
Message
Cal 1
747
solution low
warning
Cal 2
748
Solution
low warning
Cleaning
749
Solution
low warning
750
Solution
projection
adjusted
Message 751 is only found in the Activity Log to inform the user
751
about activities that have taken place.
The message is blank (empty) in the database, and when an activity
occurs the actual status information is appended to it resulting in
the logged 751-message.
If the setting "Log All Measuring Activities" is enabled in
Miscellaneous Setup, all wet section activities will be logged in the
Activity Log as 751-messages.
753
Service
Electrode
Updatings
enabled
ABL800 FLEX Operator's Manual
Interpretation
Cal 1 Solution level is
below the defined
warning level.
Cal 2 Solution level is
below the defined
warning level.
Cleaning Solution level
is below the defined
warning level.
The estimated solution
level was adjusted by
the operator.
Logging of sensor
signals was enabled.
This may be due to
ongoing technical
service activities on the
analyzer. The condition
does not affect analyzer
functions.
Operator action
Replace container as
required.
Removal condition:
Solution level above
warning level.
Replace container as
required.
Removal condition:
Solution level above
warning level.
Replace container as
required.
Removal condition:
Solution level above
warning level.
No action required.
Contact Radiometer service
representative if logging
was not enabled on
purpose.
Removal condition:
Electrode signal logging
disabled by Radiometer
service representative.
Continued on next page