Miscellaneous setup,
List of options
(continued)
Log all
measurement
activities
Auto temp. unit
conversion
Show screen saver
Show parameter bar If disabled, the parameter bar will not be shown on the
To activate/deactivate an option, do the following:
Step
Selecting HbF
This option disables HbF correction for all levels, or enables it for all levels or for
correction
HbF levels higher than 20 % (HbF correction will be automatically disabled if
FCOHb < −0.5 % in order to decrease a false HbF detection on adult blood
option
samples and to ensure non-biased oximetry results).
The guidelines for selecting/deselecting HbF correction are as follows:
For neonatal samples: Use "Enabled for all levels".
For adult samples:
NOTICE: When an adult sample is measured with HbF correction "Enabled for all
levels" or "Enabled for levels > 20 %", it will slightly affect measurement of sO
FO
samples reported with HbF present.
If this is not acceptable, the HbF correction for adult samples can be turned off
before a measurement – see Selecting HbF Correction in the Analysis Setup in this
chapter.
To select the desired option, use the arrow buttons in the box.
Continued
Option
If enabled, "Ready", "Rinse", "Aspirating" and
"Measurement" will be registered in the Activity log.
Otherwise these activities will not be registered in the
Activity log. This option aims to avoid too many entries
in the Activity log.
°C will be automatically changed to °F if the entered
temperature is over the value of 45.
The screen saver will show up if the analyzer was idle for
10 minutes.
Ready screen.
Action
1.
Scroll the list of options with the up/down arrow keys.
2.
Highlight the option and press the check button ( ) beside it. To
deactivate the option, press the check button again.
Press Close to confirm the settings and return to the Ready screen.
3.
It is important to enable HbF correction in order to
obtain correct results for ctBil, sO
FCOHb and FHHb.
Use "Disabled" or "Enabled for levels > 20 %".
Hb, FMetHb, FCOHb and FHHb, and will cause a marginal number of adult
2
3. Installation and setup
Function
, FO
Hb, FMetHb,
2
2
Continued on next page
,
2
3-87