RADIOMETER ABL800 FLEX Operator's Manual page 176

Hide thumbs Also See for ABL800 FLEX:
Table of Contents

Advertisement

4. Sample measurements
Patient result,
Parameter
status
(continued)
"....." instead of
∗ next to the value
Low
parameter
value
Patient result
Result approval is possible if the "Enable Result Approval" check button has been
approval
activated ( ) in the Miscellaneous Setup program.
To use this function, do the following:
Step
4-22
Continued
Marking
Parameter value is outside the upper or lower critical
limit (the limits outside of which a parameter value is
dangerously high or low, for the specific type of sample
being measured), but inside the reportable range.
Parameter value is outside the reportable range (user-
defined and equal to or narrower than the measuring
range).
These markings will be printed together with the result
and a verbal explanation.
A parameter cannot be calculated or exceeds the
the value
numerical limit of the analyzer.
Values with user-defined correction factors.
Critical limit
Action
1.
Press Approval on the Patient Result screen.
2.
The following buttons become available when you press the button:
Approve
Press Accept to accept the result and to send it to the
connected RADIANCE system or LIS/HIS.
Press the keyboard icon to write a note.
Press Cancel to return to the previous screen.
ABL800 FLEX Operator's Manual
Explanation
Critical limit
(no marking)
Reference Range
Reportable ange
R
High
parameter
value
Continued on next page