ABL800 FLEX Operator's Manual
Analyzer messages
Messages at
The following messages will be seen at the manager levels. The messages are
manager level
listed in numerical order.
Operator actions are listed in order of priority. Perform the first action in the list; if
unsuccessful, try the next action, etc.
No.
Message
1
Inconsistent
software
versions.
Please
contact
service.
83
Value above
reference
range
84
Value below
reference
range
85
Value is
below the
critical limit
86
Value is
above the
critical limit
Interpretation
Inconsistent software
versions for different
modules. May appear
after replacing a
complete module or as
a result of an
incomplete software
upgrade.
Parameter value is
above the user-defined
reference range.
This is only a message,
not an error.
Parameter value is
below the user-defined
reference range.
This is only a message,
not an error.
Parameter value is
below the user-defined
critical limit.
This is only a message,
not an error.
Parameter value is
above the user-defined
critical limit.
This is only a message,
not an error.
11. Troubleshooting
Operator action
Contact Radiometer service
representative.
Removal condition:
Successful software
consistency check.
No action required.
No action required.
No action required.
No action required.
Continued on next page
11-7