11. Troubleshooting
Analyzer messages,
Messages at
manager level
(continued)
11-48
Continued
No.
Message
782
RADI-
ANCE
output
queue
cleared
783
Automatic
backup
started
785
Automatic
archiving
started
786
Automatic
archiving
completed
787
Export of
data logs
started
788
Hypochlorite
detected -
remembrane
Glu, Lac, Cl
and both
Crea
electrodes
ABL800 FLEX Operator's Manual
Interpretation
The output queue was
cleared in RADIANCE
Connection Setup.
Automatic backup
(selected in Disk
Functions Setup) has
started.
Automatic archiving
(selected in Disk
Functions Setup) has
started.
Automatic archiving
(selected in Disk
Functions Setup)
completed successfully.
Export of data logs was
started by the user.
Measurement on a
substance resembling
hypochlorite has been
performed.
Any installed Glu, Lac,
Cl and Crea electrodes
have been damaged and
must be remembraned
before blood
measurements can be
performed.
As a precaution all
parameters are marked
with 'Calibration
Expired'.
Operator action
Informative message. No
action required.
Informative message. No
action required.
Informative message. No
action required.
Informative message. No
action required.
Informative message. No
action required.
- Remembrane any
installed Glu, Lac, Cl and
Crea (A and B) electrodes.
- Perform a 2-point
calibration.
Removal condition:
Successful 2-point
calibration.
Continued on next page