General information,
Recording
Step
replacements in
the Fluid
Replacements
mode
(continued)
If desired, enter the operator initials and a note for the Activity Log.
Restart after
Press Restart and then Accept on the Confirm Replacement screen.
replacements
If the barcode for the new Rinse, Calibration or Cleaning solution has not been
entered, the analyzer will not restart and the pop-up window appears with the
message to enter the solution barcode. Press Back to return to the Recording Fluid
Replacements screen and to scan the barcode.
Press Restart and then Accept. The analyzer performs a restart sequence,
depending on the replaced item.
After electrode and electrode membrane replacements, calibrations are performed
more frequently during the first 24 hours – see Unscheduled Calibrations in
chapter 6 of this manual.
Continued
Action
3.
Scan the barcode on a new component, or press the keyboard icon in
the "For each item" box, type in the barcode and confirm with Enter.
The replaced items appear in the "Recorded replacements" box as
soon as the code has been scanned or typed.
4.
Install a new component according to recommended procedure.
5.
Repeat steps 1-4 for another replacement in a similar manner.
6.
To remove an item from the recorded replacements list, highlight it
with the up/down arrow buttons in the box and press Undo.
7. Replacements
7-5