11. Troubleshooting
Analyzer messages,
Messages at
manager level
(continued)
11-8
Continued
No.
Message
89
Measured
QC value
above the
control
range
90
Measured
QC value
below
control
range
93
Value
above the
reportable
range
94
Value below
the
reportable
range
117
LIS/HIS:
Invalid
connection
configura-
tion
ABL800 FLEX Operator's Manual
Interpretation
Measured parameter
value is above the
control range.
Measured parameter
value is below the
control range.
Parameter value is
above the reportable
range. See chapter 13:
Specifications in the
Operator's Manual
Parameter value below
the reportable range.
See chapter 13:
Specifications in the
Operator's Manual.
The comunication
configuration or the
protocol definition was
invalid.
Operator action
- Verify procedure and
repeat measurement.
- Refer to Quality Control
Systems Reference
Manual.
Crea:
- Mount a new Cleaning
Met II Solution.
- Remembrane both Crea A
and Crea B electrodes
- Verify procedure and
repeat measurement.
- Refer to Quality Control
Systems Reference
Manual.
Crea:
- Mount a new Cleaning
Met II Solution.
- Remembrane both Crea A
and Crea B electrodes
- Check for and remedy
other errors related to the
result, system messages, or
calibration status.
- Perform QC. If the QC
result is accepted, the
blood sample may be
suspected.
- Repeat measurement on a
new blood sample.
- Perform the Electrode
Troubleshooting procedure.
Check the communication
parameters specified in
Communications setup.
Continued on next page