ABL800 FLEX Operator's Manual
Radiometer default settings,
Calibration
Default settings for Calibration Schedule are as follows:
Setup defaults
(continued)
2 Point Cal
1 Point Cal
1 Point pH/BG calibration
(for the USA only)
tHb Cal
Cleaning
Start time for 1- and 2-point calibration and
cleaning
Activity after measurement
Quality Control
Default settings for Quality Control Statistics are as follows:
Setup defaults
Statistics factor
Remind to print statistics each month
Remind to export WDC data each month
Default settings for Quality Control Mandatory Input are as follows:
Mandatory temperature
Default temperature
 QC Schedule settings are deleted.
 AutoCheck Setup: replacement warning before carousel empty: No.
 Westgard Rules: all rules are 'Off' and the use of Westgard Rules is disabled.
 Rilibäk Ranges: the use of Rilibäk rules is disabled.
Continued
Activity
Item
Item
No
25
Default setting
8 hours
4 hours
2 hours
3 months
8 hours
00:00
None
Default setting
1.5
No
No
Default Setting
o
C
15. Radiometer settings
Continued on next page
15-7