3. Installation and setup
Analysis setup,
Setting age
To set the age group limits, do the following:
group limits
Step
Setting
To set the reference and critical limits for each parameter do the following:
reference and
critical limits
Step
3-20
Continued
Action
1.
Press the Age groups button to display the following screen:
2.
Use the following:
• left/right arrows to choose the age group limit you want to alter
(indicated by a blue circle containing a white cross).
• up/down arrows to scroll through the list of possible age limits. As
the list is scrolled the text on the age group bar changes accordingly.
3.
Repeat step 2 for each limit to be changed.
4.
Press Back when completed to return to the Reference Ranges and
Critical Limits screen.
5.
Activate the Age group check button.
To deactivate age group differentiation, press the check button again.
Action
1.
Highlight a parameter by using the up/down arrow buttons.
2.
Choose the sample type, sex, age group and press the Edit button.
ABL800 FLEX Operator's Manual
0 days-1 month,
1-5 months,
5-8 months,
>8 months"
Continued on next page