3. Installation and setup
Quality control setup,
QC solutions
To assign a control solution to a slot, do the following:
(continued)
Step
The following quality control solutions can be used with the ABL800 FLEX
analyzers:
QUALICHECK5+; AutoCheck5+
QUALICHECK3+; QUALICHECK5+; AutoCheck3+;
AutoCheck5+
QUALICHECK5+; AutoCheck5+
AutoCheck6+, S8377 Cleaning Met II Solution
For the solution types for each quality control system, please refer to chapter 14:
Ordering information in this manual.
You can also use a suitable non-Radiometer quality control system for your
analyzer.
3-36
Continued
Action
Highlight a slot, using the arrow buttons.
1.
2.
For control solutions from Radiometer, scan the bar code or press the
keypad button
3.
For non-Radiometer control solutions, press Add Non-R-.
Press Accept to continue with the installation of the control solution.
Press Cancel to cancel installation of the control solution.
4.
To delete a control solution, highlight the desired slot and press
Delete. The warning as in step 3 above appears.
Press Accept to continue with the removal of the control solution.
Press Cancel to cancel.
5.
To define whether an AutoCheck quality control measurement is to be
run manually or automatically, press the Manual/Auto button. Useful
if the AutoCheck module is out of order.
Quality control system
ABL800 FLEX Operator's Manual
to enter the barcode.
The warning
Analyzer
ABL805
ABL830/20/10
ABL835/25/15
ABL837/27/17
Continued on next page