11. Troubleshooting
Analyzer messages,
Messages at
manager level
(continued)
11-38
Continued
No.
Message
695
ABL discon-
nected from
RADIANCE
696
ABL<>
RADIANCE
communi-
cation error
697
Requested
AutoCheck
QC ampoule
not present in
carousel
698
Performing
AutoCheck
on last
ampoule in
schedule
instead of
non-present
699
AutoCheck
QC measure-
ment started
due to cali-
bration error
700
Scheduled
AutoCheck
not run due
to errors in
last
calibration
703
QC expired
704
AutoCheck
QC measure-
ment is
repeated
ABL800 FLEX Operator's Manual
Interpretation
The analyzer was
disconnected from
RADIANCE.
Communication error
between the analyzer
and RADIANCE.
A scheduled QC
ampoule was absent in
the carousel.
If the scheduled QC
ampoule was absent in
the carousel, the
analyzer performs QC
measurement on the
ampoule scheduled
previously.
The analyzer was set up
to perform AutoCheck
measurements in case of
1- or 2-point calibration
errors.
Last calibration
contained an error, and
the analyzer was set up
to suspend AutoCheck
measurements in case
of calibration errors.
QC measurement is
overdue (corrective
action "Lock analyzer"
has been selected in
Setup program
Corrective Actions).
The scheduled QC
measurement was not
accepted; the
measurement was
repeated as requested in
Setup program
Corrective Actions.
Operator action
No action required.
Contact Radiometer service
representative.
Refill AutoCheck carousel
according to the packing
list.
Refill AutoCheck carousel
according to the packing
list.
Check Calibration Status
and remedy any reported
calibration errors.
Check Calibration Status
and remedy calibration
errors.
Perform a quality control
measurement.
Removal condition: No QC
measurements are pending.
No action required.
Continued on next page