RADIOMETER ABL800 FLEX Operator's Manual page 305

Hide thumbs Also See for ABL800 FLEX:
Table of Contents

Advertisement

ABL800 FLEX Operator's Manual
Analyzer messages,
Messages at
manager level
(continued)
Continued
No.
Message
505
Met II upper
liquid
sensor did
not detect
Cleaning
Solution
506
El/Met
upper liquid
sensor did
not detect
Cleaning
Solution
507
pH/BG
upper liquid
sensor did
not detect
Cleaning
solution
508
Rinse error
509
pH/BG
upper liquid
sensor did
not receive
Rinse from
above
510
El/Met
upper liquid
sensor did
not receive
Rinse from
above
Interpretation
The Met II module was
not filled properly
during Cleaning
program. The activity
is aborted.
The El/Met module
was not filled properly
during Cleaning
program. The activity
is aborted.
The pH/BG module
was not filled properly
during Cleaning
program. Cleaning is
aborted.
Rinse Solution was not
detected during Rinse
program.
Measurements and
calibrations cannot be
performed until the
condition is removed.
The pH/BG module
was not filled properly
during Rinse program;
automatic Refill is
attempted.
The El/Met module
was not filled properly
during Rinse program;
automatic Refill is
attempted.
11. Troubleshooting
Operator action
- Check the Cleaning
solution for sufficient
volume. Replace as
required.
- Perform Fluid transport
troubleshooting procedure.
Removal condition.
Successful Cleaning.
- Check the rinse solution
for sufficient volume.
Replace as required.
- If the message "Inlet flow
impeded" (765) has been
reported, replace or clean
inlet probe.
- Perform Fluid transport
troubleshooting procedure.
Removal condition:
Successful Rinse.
- Check the Rinse Solution
for sufficient volume.
Replace as required.
- Perform Fluid transport
troubleshooting procedure.
Removal condition:
Successful Rinse.
Continued on next page
11-23