RADIOMETER ABL800 FLEX Operator's Manual page 345

Hide thumbs Also See for ABL800 FLEX:
Table of Contents

Advertisement

ABL800 FLEX Operator's Manual
Analyzer messages,
Messages at
manager level
(continued)
Continued
No.
Message
931
No sample
in Met II's
lower liquid
sensor
within time
limit
932
No sample
in Met II's
upper liquid
sensor
within time
limit
934
Inlet not in
CONE3
position
before
timeout
prior to
wash with
MetCal1
935
Cal1
Solution -
Crea will
expire soon
936
Cal2
Solution -
Crea will
expire soon
Interpretation
The sample did not fill
Met II module
properly. The activity
is aborted.
Inlet probe was not in
proper position within
the required time limit.
The additive used in
the Calibration
Solution 1 to calibrate
Crea will expire soon.
This is a warning - the
solution lifetime is 14
days.
The additive used in
the Calibration
Solution 2 to calibrate
Crea will expire soon.
This is a warning - the
solution lifetime is 14
days.
11. Troubleshooting
Operator action
- Check the sample for
clots.
- Ensure that sample
volume matches selected
sample mode.
- Perform the Fluid
Transport Troubleshooting
procedure.
- Remove the inlet gasket
and inspect the condition of
inlet probe and gasket.
Replace as necessary.
Install all parts.
- Restart analyzer.
- Contact Radiometer
Service Representative.
Removal condition:
Successful inlet calibration.
Replace the solution before
the lifetime is actually
exceeded.
Replace the solution before
the lifetime is actually
exceeded.
Continued on next page
11-63